Skip to main content
. Author manuscript; available in PMC: 2011 Jul 6.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2009 May;56(5):995–1005. doi: 10.1109/tuffc.2009.1131

TABLE I.

Ultrasound Treatment Parameters for Each Treatment Location or Each lesion.

Each treatment location
Each lesion (volume ~0.5 cm3)
Sample size
# of pulses Treatment time (s) Delivered energy (J)* Total treatment time Total delivered energy (J)**
0 0 0 0 0 9
100 1 0.8 27 s 20.7 6
300 3 2.3 1.4 min 62.0 6
500 5 3.8 2.3 min 103.3 6
700 7 5.4 3.2 min 144.7 6
1000 10 7.7 4.5 min 206.7 9
1500 15 11.5 6.8 min 310.0 7
2000 20 15.3 9 min 413.4 6
*

The delivered energy per treatment location is calculated as ISATA × ABCS × T where ABCS is the −6 dB beam cross-sectional area measured in free field, and T is the treatment time per treatment location.

**

The total delivered energy for each lesion is calculated as (Delivered energy to each treatment location) × (Total number of treatment locations).