Skip to main content
. 2011 Jun 3;12(2):723–731. doi: 10.1208/s12249-011-9639-5

Table II.

Percent Δ8-THC Loss of in Different Containers at Two Different Concentrations

Type of container Percentage of drug loss in 30 min (0.5 μg/ml, 5%v/v ethanol) Percentage of drug loss in 30 min (0.15 μg/ml, 0.5% v/v ethanol)
Polyethylene inserts 86.4 and 76.0 (n = 2) ND
Polypropylene inserts 78.7 ± 8.1 ND
Ultra high grade Polypropylene 41.2 ± 4.7 47.0 ± 2.3
Clear glass vial (20 mL) 1.8 ± 1.6 1.7 ± 1.3
Clear glass vial (5 mL) 0.9 ± 1.1 1.9 ± 1.4
HPLC vials (1 mL) 0 1.4 ± 1.8

These studies were carried out for 30 min at room temperature. Values are represented as mean ± sd (n = 6). All glass vials used in this study met USP Type I specification

ND not detectable