Table 1.
LV functional parameters determined with manually adjusted and automatic measurements for three different software tools
Manual | Auto | P value | r value | |
---|---|---|---|---|
LVEF (%) | ||||
Vitrea | 64.3 ± 11.5 | 62.6 ± 11.9 | <0.05 | 0.93 |
Circulation | 61.8 ± 10.7 | 35.7 ± 13.6 | <0.001 | 0.45 |
Aquarius | 61.9 ± 11.3 | 61.6 ± 12.0 | 0.27 | 0.96 |
EDV (ml) | ||||
Vitrea | 155.1 ± 47.1 | 157.8 ± 40.8 | 0.15 | 0.94 |
Circulation | 170.2 ± 52.3 | 146.0 ± 50.2 | <0.001 | 0.93 |
Aquarius | 167.4 ± 51.2 | 159.5 ± 46.1 | <0.001 | 0.94 |
ESV (ml) | ||||
Vitrea | 57.7 ± 35.1 | 61.4 ± 35.2 | <0.001 | 0.98 |
Circulation | 66.5 ± 36.7 | 94.6 ± 45.5 | <0.001 | 0.92 |
Aquarius | 66.5 ± 39.8 | 58.7 ± 31.0 | <0.001 | 0.98 |
MM (g) | ||||
Vitrea | 176.9 ± 56.3 | 177.7 ± 49.3 | 0.40 | 0.94 |
Circulation | 183.2 ± 54.0 | 231.5 ± 81.1 | <0.001 | 0.86 |
Aquarius | 162.0 ± 54.3 | 190.7 ± 70.1 | <0.001 | 0.96 |
Time (s) | ||||
Vitrea | 417.7 ± 91.0 | 17.6 ± 3.2 | <0.001 | 0.37 |
Circulation | 352.8 ± 67.5 | 61.3 ± 20.4 | <0.001 | 0.22 |
Aquarius | 213.9 ± 32.6 | 74.4 ± 8.9 | <0.001 | 0.20 |
Data are presented as mean ± SD