TABLE III.

Top ten simulated designs by reflection coefficient, with relevant performance data. Diameter, length, and aspect ratio are measured from the SAR = 30 W∕kg isocontour. Many of these designs produced very elongated heating patterns that would not be suitable for clinical use.

Rank G1 (mm) L (mm) G2 (mm) Objective S11 (dB) Diameter (mm) Length (mm) Aspect Ratio
1 8 20 10 0.999 −30.7 14.5 71.9 0.20
2 1 8 3 0.644 −27.8 19.8 30.1 0.66
3 2 8 2 0.947 −27.2 19.3 38.1 0.51
4 7 20 10 1.000 −25.3 14.2 72.7 0.20
5 1 7 2 0.853 −25.0 19.5 33.5 0.58
6 8 19 10 1.000 −24.3 14.4 71.9 0.20
7 9 20 9 1.000 −24.1 14.5 71.3 0.20
8 9 19 10 1.000 −23.8 14.8 71.3 0.21
9 8 20 9 1.001 −22.8 14.3 72.5 0.20
10 1 8 2 0.827 −22.6 18.9 31.7 0.60