Skip to main content
. 2011 Jun 28;38(7):4232–4240. doi: 10.1118/1.3601019

TABLE IV.

Top ten simulated designs by heating zone aspect ratio, with relevant performance data. Diameter, length, and aspect ratio are measured from the SAR = 30 W∕kg isocontour. Several of these designs create reflection coefficients that would increase cable shaft heating and degrade performance in tissue.

Rank G1 (mm) L (mm) G2 (mm) Objective S11 (dB) Diameter (mm) Length (mm) Aspect Ratio
1 1 7 7 0.794 −9.0 20.4 25.4 0.80
2 1 7 6 0.614 −10.7 20.6 26.0 0.79
3 1 7 5 0.552 −12.9 20.1 27.3 0.74
4 1 7 8 1.254 −7.7 18.1 25.7 0.70
5 1 8 4 0.520 −20.9 20.4 29.2 0.70
6 1 6 8 1.369 −6.9 18.3 26.6 0.69
7 1 6 9 1.423 −6.1 17.9 26.0 0.69
8 1 7 4 0.612 −15.7 19.9 29.1 0.68
9 1 6 10 1.524 −5.4 17.6 26.3 0.67
10 1 8 3 0.644 −27.8 19.8 30.1 0.66