TABLE II.

Measurement results in two MRI scanners. Values are average of five repetitions.

  1.5 T scanner 3.0 T scanner
SNR    
 Phantom center 20.7 60.2
  Uncertainty 2.1 2.5
 Phantom periphery    
  Median (minimum, maximum) 37.9 (34.0, 42.0) 106.7 (87.2, 125.7)
  Uncertainty 3.0 13.3
EPI readout distortion along phase-encoding direction (mm)    
 Median (minimum, maximum) −4.6 (−6.9, −0.4) −7.1 (−10.0, −0.5)
 Uncertainty 0.4 0.6
Image distortion from diffusion-encoding gradient in  phase-encoding direction (mm)    
 Median (minimum, maximum)    
  yb, before registration 0.06 (0.01, 0.16) 0.09 (−0.20, 0.41)
  σ, before registration 0.49 (0.27, 0.77) 0.40 (0.08, 0.69)
  yb, after registration 0.04 (−0.04, 0.11) 0.00 (−0.20, 0.11)
  σ, after registration 0.58 (0.41, 0.81) 0.33 (0.08, 0.53)
  uncertainty of yb 0.08 0.06
  uncertainty of σ 0.13 0.13
FA    
 Pixel-based, mean ± SD 0.098 ± 0.037 0.044 ± 0.017
 Uncertainties in mean ± SD 0.001 ± 0.001 0.001 ± 0.001
 ROI-based 0.0083 0.0098
 Uncertainty 0.0008 0.0007
MD (10−3 mm2/s)    
 Pixel-based, mean ± SD 2.025 ± 0.065a 2.038 ± 0.039b
 Uncertainties in mean ± SD 0.001 ± 0.002 0.001 ± 0.002
 ROI-based 1.998 2.033
 Uncertainty 0.001 0.002
a

Temperature = 19.3 °C, reference MD value 1.988 × 10−3 mm2/s.

b

Temperature = 19.4 °C, reference MD value 1.993 × 10−3 mm2/s.