Table 2.

Opertor 2 results.

Phantom configuration True sphere/
background ratio
SPECT alone SPECT-CT
Isocontour (%) Measured volume (cm3) Error (%) Isocontour (%) Measured volume (cm3) Error (%)
18,5 MBq/sphere Cylinder 1 1% 7087 5,52 2% 6521 −2,90
Sphere 1 2.8 31% 207 276,36 48% 57,2 4,00
Sphere 2 7.6 13% 104,5 397,62 28% 24,4 16,19
Sphere 3 9.6 19% 27,4 71,25 29% 16,9 5,62
Sphere 4 19.6 8% 29,2 265,00 19% 10,9 36,25

37 MBq/sphere Cylinder 1 2% 6180 −7,98 1% 6700 −0,24
Sphere 1 6 7% 169,7 208,55 34% 53,56 −2,62
Sphere 2 17 3% 113,7 441,43 23% 24,34 15,90
Sphere 3 23.4 3% 119,78 648,63 27% 14,9 −6,88
Sphere 4 43.8 4% 53,48 568,50 33% 9,8 22,50

55,5 MBq/sphere Cylinder 1 1% 6737 0,31 1% 6737 0,31
Sphere 1 9.9 10% 159,3 189,64 35% 55,7 1,27
Sphere 2 29.6 4% 85 304,76 25% 22,7 8,10
Sphere 3 38.6 7% 51,5 221,88 36% 15,5 −3,13
Sphere 4 73.1 6% 44,1 451,25 36% 12 50,00

74 MBq/sphere Cylinder 1 1% 6565 −2,10 1 6565 −2,10
Sphere 1 16 15% 99,5 80,91 32% 56 1,82
Sphere 2 45 6% 54,1 157,62 23% 23,7 12,86
Sphere 3 57 5% 42,8 167,50 25% 17,4 8,75
Sphere 4 114 2% 35,8 347,50 20% 8,7 8,75

55 MBq Cylinder 2 17% 887 14,60 27% 801 3,49
116 MBq Cylinder 2 19% 855 10,47 29% 775 0,13
72 MBq Cylinder 3 21% 520 9,94 27% 481 1,69