. 2011 Jul;49(7):2528–2532. doi: 10.1128/JCM.00339-11

Table 1.

Limit of detection and reproducibility of MALDI Biotyper system for yeast identification directly from a positive blood culture specimen

Sample 2-4 dilution No. of CFU/ml Identification Score
% CV
Specimen 1 Specimen 2 Specimen 3 Mean ± SD
Undiluted 5.9 × 106 Reliable identification 2.230 2.112 2.211 2.184 ± 0.063 2.90
2.234 2.245 2.170 2.216 ± 0.041 1.83
2.204 2.354 2.229 2.262 ± 0.080 3.55
1:10 5.9 × 105 Reliable identification 1.985 2.087 2.013 2.028 ± 0.053 2.60
2.160 2.057 2.079 2.099 ± 0.054 2.58
1.942 1.984 2.006 1.977 ± 0.033 1.64
1:100 5.9 × 104 Not reliable identification 1.328 1.230 1.276 1.278 ± 0.049 3.84
1.174 1.381 1.273 1.276 ± 0.104 8.11
1.382 1.361 1.370 1.371 ± 0.011 0.77
1:1,000 5.9 × 103 Not reliable identification 1.446 1.540 1.483 1.489 ± 0.047 3.18
1.231 1.264 1.245 1.246 ± 0.017 1.33
1.212 1.363 1.288 1.288 ± 0.076 5.86