Table 2. Effect of temperature and humidity on performance of the POC test compared to Amplicor PCR.
Temperature threshold: 31.4°C | ||||
POC test – reader 1 | POC test – reader 2 | |||
Below threshold | Above threshold | Below threshold | Above threshold | |
True positive | 10 | 9 | 10 | 7 |
True negative | 956 | 505 | 961 | 531 |
False positive | 21 | 100 | 16 | 74 |
False negative | 7 | 2 | 7 | 4 |
Sensitivity | 58.8(32.9–81.6) | 81.8(48.2–97.7) | 58.8(32.9–81.6) | 63.6(30.8–89.1) |
Specificity | 97.9(96.7–98.7) | 83.5(80.3–86.4) | 98.4(97.4–99.1) | 87.8(84.9–90.3) |
PPV | 32.3(16.7–51.4) | 8.3(3.8–15.1) | 38.5(20.2–59.4) | 8.6(3.5–17.0) |
NPV | 99.3(98.5–99.7) | 99.6(98.6–100) | 99.3(98.5–99.7) | 99.3(98.1–99.8) |
Values in parentheses are 95% CI.