Table 6.
Individual data for infusion verification and validation for the 30 μl/h Valsartan® infusion.
Animal number | Filled | Programmed | Remaining (theoretical) | Remaining (extracted) | Volume administered (measured) | Difference (volume administered) | #x00025; Difference (volume administered) | |
---|---|---|---|---|---|---|---|---|
201001586 | 900 | 755.4 | 144.6 | 170 | 730 | −25.4 | −3.36 | |
201001587 | 900 | 755.4 | 144.6 | 140 | 760 | 4.6 | 0.61 | |
201001588 | 900 | 755 | 145 | 150 | 750 | −5 | −0.66 | |
201002026 | 900 | 756 | 144 | 120 | 780 | 24 | 3.17 | |
201002027 | 900 | 756 | 144 | 180 | 720 | −36 | −4.76 |