Protein sample /Technical parameters | GluFibrino-peptidemonomer1.6 kDa | Myoglobinmonomer17 kDa | Hemoglobintetramer67 kDa | Transferrinmonomer80 kDa | GroEL14-mer801 kDa |
Backing pressure, mBar | 4.4 | 5.0 | 5.1 | 5.1 | 6.5 |
Trap pressure, mBar | 1.6x10-2 | 2.4x10-2 | 2.4x10-2 | 2.6x10-2 | 2.8x10-2 |
IMS Pressure, mBar | 4.4x10-1 | 4.4x10-1 | 4.4x10-1 | 4.4x10-1 | 4.2x10-1 |
Sampling cone voltage, V | 46 | 80 | 80 | 80 | 118 |
Extraction cone voltage, V | 1.7 | 1 | 1 | 1 | 3 |
Bias voltage, V | 20 | 20 | 25 | 25 | 50 |
Trap collision energy, V | 20 | 15 | 15 | 15 | 80 |
Transfer collision energy, V | 5 | 12 | 12 | 12 | 15 |