Table 2. Validation results of PPD in human urine.

Intra- and interday Accuracy and Precision in human urine
nominal conc. [µmol/L] mean calc. conc.[µmol/L] Accuracy (bias %) Precision (% CV)
MALDI-MS/MS / HPLC-UV MALDI-MS/MS / HPLC-UV
Intra-day (n = 3) 1000 960 −3.9 2.7 24.1 10.5
750 745 −0.6 −9.4 9.9 5.2
500 512 2.5 2.2 8.9 8.8
250 261 4.4 9.2 13.2 1.8
150 157 4.5 4.7 2.4 9.4
100 91 −8.9 n.d. 9.5 n.d.
50 51 2.1 n.d. 16.2 n.d.

Inter- and intraday accuracy and precision measurements (n = 3 for each value). The mean relative error of each concentration is expressed as % bias and the reproducibility is depicted as the coefficient of variation (% CV). Intraday accuracy and precision were also determined on the HPLC-UV system.