Technique settings and dosimetry for task-specific protocols in the thoracic and lumbar spine. “Bone Protocol” refers to a lower dose technique sufficient for high-contrast bone visualization. “Soft-Tissue Protocol” refers to a higher dose technique sufficient for soft-tissue visualization.

  Bone protocol Soft-tissue protocol
  Thoracic spine Lumbar spine Thoracic spine Lumbar spine (LoRes) Lumbar spine (HiRes)
KVp 100 100 100 100 120
mA 0.6 1.3 2.3 2.3 2.3
mAs 60 130 230 230 460
# Projections 200 200 200 200 400
Dscan (mGy) 1.81 3.16 4.26 5.1 10.6
Dw (mGy) 1.91 3.70 4.61 6.01 12.50