Table 2.

In-air exposure, exposure rate, effective dose, and effective dose rate, assuming standard shielding apparel (Pb apron etc.) at locations about the table and throughout the operating room. The exposure (mR/mGy) and effective dose (mSv/mGy) are per unit absolute dose from a single CBCT scan. The corresponding exposure rate (mR/min) and effective dose rate (mSv/min) correspond to the rate during the scan.

Position Xshielded (mR/mGy) Eshielded (mSv/mGy) X·shielded (mR/min) E·shielded (mR/min)
P1 – 57 cm 2.38 0.021 2.23 0.020
P2 – 43 cm 2.48 0.022 2.33 0.021
P3 – 57 cm 3.61 0.033 3.38 0.031
P4 – 70 cm 1.90 0.017 1.78 0.016
P5 – 90 cm 0.49 0.004 0.46 0.004
P6 – 70 cm 1.40 0.013 1.31 0.012
P7 – 110 cm 0.36 0.003 0.34 0.003
1.0 m to isocenter 0.79 0.007 0.74 0.007
1.5 m to isocenter 0.13 0.001 0.12 0.001
2.0 m to isocenter 0.01 0.0001 0.01 0.0001