Table 2.
Hardware Characteristics of the Three Small-Animal MRI Scanners.
| Hardware Component | System #1 | System #2 | System #3 |
|---|---|---|---|
| Magnet | 4.7 T, Oxford Mk I (i.d., 40.0 cm) | 4.7 T, Oxford Mk II (i.d., 33.0 cm) | 11.7 T, Magnex (i.d., 26.0 cm) |
| MRI Console | Varian DirectDrive™ System | Varian DirectDrive™ System | Varian UnityINOVA™ system |
| Gradient Coil i.d.a | 21 cm | 12 cm | 8 cm |
| Gmax | 28 G/cm | 58 G/cm | 120 G/cm |
| trise (0 – Gmax) | 650 μs | 270 μs | 298 μs |
| Imax | 300 A | 300 A | 200 A |
| slew rate (dG/dt) | 0.043 G/cm·μs | 0.215 G/cm·μs | 0.403 G/cm·μs |
| RF Coil | linear Litz volume coil, 73 mm i.d. | linear Litz volume coil, 73 mm i.d. | quadrature Litzcage coil, 38 mm i.d. |
All of these gradient sets were manufactured by Magnex Scientific (now a part of Agilent Technologies) and are water-cooled, actively-shielded, high-duty type coils.