Skip to main content
. Author manuscript; available in PMC: 2012 Sep 1.
Published in final edited form as: J Magn Reson. 2011 Jun 25;212(1):116–123. doi: 10.1016/j.jmr.2011.06.016

Table 2.

Hardware Characteristics of the Three Small-Animal MRI Scanners.

Hardware Component System #1 System #2 System #3
Magnet 4.7 T, Oxford Mk I (i.d., 40.0 cm) 4.7 T, Oxford Mk II (i.d., 33.0 cm) 11.7 T, Magnex (i.d., 26.0 cm)
MRI Console Varian DirectDrive System Varian DirectDrive System Varian UnityINOVA system
Gradient Coil i.d.a 21 cm 12 cm 8 cm
Gmax 28 G/cm 58 G/cm 120 G/cm
trise (0 – Gmax) 650 μs 270 μs 298 μs
Imax 300 A 300 A 200 A
slew rate (dG/dt) 0.043 G/cm·μs 0.215 G/cm·μs 0.403 G/cm·μs
RF Coil linear Litz volume coil, 73 mm i.d. linear Litz volume coil, 73 mm i.d. quadrature Litzcage coil, 38 mm i.d.
a

All of these gradient sets were manufactured by Magnex Scientific (now a part of Agilent Technologies) and are water-cooled, actively-shielded, high-duty type coils.