Table 5.

Calculated specificity and sensitivity of the Seeplex DV assay before discordant analysis

Result by Seeplex DV No. of specimens with PHL test result
% sensitivitya % specificityb
EM
rRT-PCR
True False True False
Adenovirus
    Positive 28 1
    Negative 1 168
        Total 29 169 97 (28/29) 99.4 (168/169)
Rotavirus
    Positive 51 1
    Negative 1 145
        Total 52 146 98 (51/52) 99.3 (145/146)
Norovirus GI
    Positive 18 1
    Negative 1 178
        Total 19 179 95 (18/19) 99.4 (178/179)
Norovirus GII
    Positive 37 1
    Negative 1 159
        Total 38 160 97 (37/38) 99.4 (159/160)
a

Numbers in parentheses are no. positive by Seeplex DV/total no. “true” by PHL test.

b

Numbers in parentheses are no. negative by Seeplex DV/total no. “false” by PHL test.