Skip to main content
. 2011 Sep;49(9):3154–3162. doi: 10.1128/JCM.00599-11

Table 6.

Calculated specificity and sensitivity of the Seeplex DV assay after discordant analysis

Assay, virus, and result No. of specimens with Seeplex DV result
% sensitivitya % specificityb
True False
Home brew multiplex rRT-PCR
    Adenovirus
        Positive 28 1
        Negative 1 168
            Total 29 169 97 (28/29) 99.4 (168/169)
    Rotavirus
        Positive 51 0
        Negative 0 146
            Total 51 146 100 (51/51) 100 (146/146)
RealStar
    Norovirus GI
        Positive 18 0
        Negative 0 179
            Total 18 179 100 (18/18) 100 (179/179)
    Norovirus GII
        Positive 38 0
        Negative 0 159
            Total 38 159 100 (38/38) 100 (159/159)
a

Numbers in parentheses are no. positive by reference assay/total no. “true” by Seeplex DV.

b

Numbers in parentheses are no. negative by reference assay/total no. “false” by Seeplex DV.