Table 3. Correlation coefficients, mean difference (MD) with 95% confidence intervals (95% CIs), and limits of agreement (LoA) for retinal nerve fibre layer (RNFL) measurements of the optic nerve head obtained with Stratus OCT and Spectralis HRA+OCT.
Region | Correlation | MD (μm) (95% CI) | LoA (μm) |
---|---|---|---|
Average | 0.73 | 6.2 (4.3, 8.1) | −4.1, 16.5 |
Superior | 0.87 | 8.5 (6.1, 11) | −4.7, 21.8 |
Inferior | 0.84 | 8.1 (5.3, 10.9) | −7.1, 23.3 |
Nasal | 0.74 | 10.3 (6.9, 13.7) | −7.8, 28.4 |
Temporal | 0.61 | 22.6 (19.3, 25.9) | 4.7, 40.5 |