Table 3. Correlation coefficients, mean difference (MD) with 95% confidence intervals (95% CIs), and limits of agreement (LoA) for retinal nerve fibre layer (RNFL) measurements of the optic nerve head obtained with Stratus OCT and Spectralis HRA+OCT.

Region Correlation MD (μm) (95% CI) LoA (μm)
Average 0.73 6.2 (4.3, 8.1) −4.1, 16.5
Superior 0.87 8.5 (6.1, 11) −4.7, 21.8
Inferior 0.84 8.1 (5.3, 10.9) −7.1, 23.3
Nasal 0.74 10.3 (6.9, 13.7) −7.8, 28.4
Temporal 0.61 22.6 (19.3, 25.9) 4.7, 40.5