Skip to main content
. 2011 Mar 25;25(5):642–650. doi: 10.1038/eye.2011.60

Table 2. Intraocular pressure measurements results.

  GAT IOP iCare IOP P^ r* GAT-iCare
  (mm Hg) (mm Hg)   P (mm Hg)
Control group (n=101) 17.6±4.9 (9.5–28) 14.9±4.5 (8–24.4) <0.0001 0.83 2.7±3.4 (−5/9)
        <0.0001  
Healthy subjects (n=58) 15.3±3.8 (9.5–21) 13.9±4.1 (8–20) <0.0001 0.86 1.4±3.5 (−5/9)
        <0.0001  
POAG patients (n=43) 20.6±4.5 (12–28) 16.2±4.7 (10–24.4) <0.0001 0.82 4.4±2.4 (0/9)
        <0.0001  
PK group (n=34) 20.3±5.0 (12–34) 14.7±5.4 (8–31) <0.0001 0.76 5.5±3.6 (−3/12)
        <0.0001  
ALTK group (n=20) 14.1±4.6 (6–24) 11.8±3.7 (5–18) <0.0001 0.78 2.3±2.9 (−4/7)
        <0.0001  
DSAEK group (n=19) 17.7±3.8 (11–25) 13.6±4.6 (8–26) <0.0001 0.70 4.0±3.3 (−2/10)
        <0.001  
Edematous grafts group (n=17) 9.1±3.1 (5–14) 13.9±3.8 (10–22) <0.0001 0.68 −6.5±1.9 (−11/−4)
        <0.002  
P** <0.0001 <0.01     <0.0001

Abbreviations: ALTK, automated-lamellar-therapeutic keratoplasty; DSAEK, Descemet-stripping-automated-endothelial keratoplasty; PK, penetrating keratoplasty; POAG, primary open-angle glaucoma; ^, paired t-test

Results are given as mean±SD (95% confidence interval). *Pearson's correlation coefficient; **analysis of variance.