Table 1.

Summary of terms used in cascaded systems analysis of the proposed extremities CBCT scanner. Nominal values were used in the experiments unless mentioned otherwise.

Term Definition Nominal value
MTFFPD(fu,fv) Modulation transfer function for 2D detector cascade
MTFspot(fu,fv) Modulation transfer function associated with focal spot blur
SQ(fu,fv) NPS associated with x-ray quantum noise in the 2D projection
SE(fu,fv) NPS associated with electronic noise in the 2D projection
(fu,fv) Frequency components of the detector Fourier domain (expressed in the detector plane)
(fu′,fv′) Frequency components of the detector Fourier domain (expressed in the imaging plane)
qo Mean fluence incident on the detector for a given projection (nominal beam: 90 kVp + 0.3 mm Cu + 4 mm Al; knee model, detector signal=100 × additive noise) 2200 (x-rays/mm2)
m Mean detector signal 2 00 000 e/pixel
g1 Quantum detection efficiency (mean fraction of incident photons interacting in the scintillator, imaging conditions as for qo) 0.78
g2 Mean number of optical photons generated per x-ray interaction 1905
g4 Coupling efficiency of optical photons to sensor photodiode 0.6
apix Detector sampling interval 0.194 mm (full-res)
0.388 mm (half-res)    
apd Detector sensitive aperture 0.16 mm (full-res)
0.32 mm (half-res)    
σadd Additive electronics noise 2000 e/pixel
aspot Focal spot size 0.5 mm
N Number of projections acquired across circular orbit 480
SDD Source-detector distance 550 mm
SAD Source-isocenter distance 420 mm
M Magnification factor, SDD/SAD 1.3
τ X-ray pulse length 30 ms