Skip to main content
. Author manuscript; available in PMC: 2011 Sep 15.
Published in final edited form as: J Anal Toxicol. 2009 Jun;33(5):243–252. doi: 10.1093/jat/33.5.243

Table I.

Mass Spectrometric Transitions at Specific Collision Energies (CE), Retention Times (Rt), and Imprecision (%CV) after 50 Injections for Morphine, Codeine, 6-Acetylmorphine (6AM), Benzoylecgonine (BE), Cocaine, 2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), Methadone, and Deuterated IStd

Segment Compound Quantifier Transition (CE) Qualifier Transitions (CE) Rt (min) %CV Rt (n = 50)
I Morphine 286.2 → 201.1 (37) 286.2 → 229.0 (37) 4.50 0.74
286.2 → 268.1 (37)
Morphine-d6 292.3 → 201.1 (37) 4.48 0.71
II Codeine 300.2 → 215.1 (37) 300.2 → 243.0 (37) 5.37 0.3
300.2 → 282.2 (37)
Codeine-d6 306.2 → 218.1 (37) 5.35 0.31
6AM 328.2 → 211.1 (38) 328.2 → 268.1 (38) 5.66 0.25
328.2 → 193.1 (38)
6AM-d6 334.1 → 211.1 (38) 5.64 0.25
BE 290.1 → 168.1 (29) 290.1 → 168.1 → 150.0 (32) 6.10 0.20
290.1 → 168.1 → 82.1 (32)
BE-d8 298.2 → 171.1 (29) 6.07 0.20
III Cocaine 304.1 → 182.1 (30) 304.1 → 182.1 → 150.1 (32) 6.92 0.22
304.1 → 182.1 → 82.1 (32)
Cocaine-d3 307.1 → 185.1 (30) 6.91 0.24
IV EDDP 278.2 → 249.1 (40) 278.2 → 249.1 → 234.2 (34) 8.61 0.50
EDDP-d3 281.3 → 249.1 (40) 8.60 0.49
Methadone 310.1 → 265.0 (27) 310.1 → 265.0 → 247.1 (31) 8.72 0.62
310.1 → 265.0 → 219.1 (31) 8.69 0.65
Methadone-d9 319.2 → 268.1 (27)