Table III.

Imprecision and Analytical Recovery for Morphine, Codeine, 6-Acetylmorphine (6AM), Benzoylecgonine (BE), Cocaine, 2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP), and Methadone

Imprecision*
Analytical Recovery
Compound QC Intra-assay (n = 5) Interassay (n = 20) Total Intra-assay (n = 5) Interassay (n = 20)
Morphine Low 5.9 0.0 5.9 101.8 100.4
Medium 4.9 5.0 7.01 92.9 100.7
High 7.9 4.1 8.9 92.8 97.1
Codeine Low 8.7 4.1 9.6 111.7 104.2
Medium 7.0 3.6 7.9 113.3 106.6
High 5.7 6.8 8.8 112.1 103.5
6-AM Low 7.8 7.2 10.6 89.9 98.1
Medium 4.5 0.0 4.5 111.5 110.2
High 8.2 4.0 9.1 93.1 100.3
BE Low 8.6 5.1 10.0 92.4 97.9
Medium 6.2 6.7 9.1 110.7 101.2
High 8.1 4.7 9.4 92.3 97.6
Cocaine Low 5.6 3.0 6.3 110.1 105.0
Medium 2.7 0.0 2.7 111.6 107.1
High 4.7 4.1 6.2 109.7 102.7
EDDP Low 5.3 9.1 10.5 87.5 94.5
Medium 4.2 5.3 6.8 87.7 92.6
High 5.0 9.8 11.1 84.4 92.2
Methadone Low 5.7 1.8 5.9 95.7 99.1
Medium 6.5 5.8 8.7 92.7 96.5
High 4.9 6.3 8.0 112.1 100.2
*

Imprecision is expressed as %CV.

Analytical recovery is expressed as % of target concentration.