Table 2.

Resolution integral, characteristic resolution and depth of field measurements for each of the transducers

Scanner Transducers Resolution integral, R Characteristic resolution, DR (μm) Depth of field, LR (mm)
Vevo 770 (tracking) RMV 704 72 93 6.7
Vevo 770 (mean, non-tracking) RMV 704 25 132 3.3
Vevo 2100 MS550S 56 188 10.5
Vevo 2100 MS550D 55 197 10.9
Diasus L10-22 44 520 23
iU22 L15-7io 72 600 43
Diasus L8-16 55 600 33
iU22 L17-5 78 630 49
Vivid 5 i13Lv 35 720 25
MicroMaxx HFL38/13-6 57 840 48

Vevo 770 (mean, non-tracking) was the mean of three sets of measurements taken using the same Edinburgh Pipe Phantom over a period of 15 months. Vevo 770 (tracking) was calculated by tracking the vertical extent of each pipe through the phantom to simulate preclinical use. The results are ranked with respect to characteristic resolution such that those transducers with smallest characteristic resolution are at the top of the table.