. 2011 Mar;37(3-2):493–501. doi: 10.1016/j.ultrasmedbio.2010.11.010

Table 3.

Resolution integral, characteristic resolution and depth of field for each of the transducers measured over a 0–10 mm depth range

Scanner Transducers Resolution integral, R Characteristic resolution, DR (μm) Depth of field, LR (mm)
Vevo 770 (tracking) RMV 704 72 93 6.7
Vevo 770 (mean, non-tracking) RMV 704 25 132 3.3
Vevo 2100 MS550S 51 181 9.2
Vevo 2100 MS550D 47 181 8.5
Diasus L8-16 17 480 8.2
Diasus L10-22 15 490 7.5
iU22 L15-7io 16 530 8.5
Vivid 5 i13Lv 15 560 8.3
iU22 L17-5 12 600 7.1
MicroMaxx HFL38/13-6 7 930 6.8

The results are ranked with respect to characteristic resolution such that those transducers with smallest characteristic resolution are at the top of the table.