Table 2.
Assaya | Detection with Hib genome copy numberb: |
|||||||
---|---|---|---|---|---|---|---|---|
106 | 105 | 104 | 103 | 102 | 10 | 1 | 0 | |
PCR | + | + | + | + | – | – | – | – |
LAMP (35 min) | + | + | + | + | + | + | ± | – |
LAMP (60 min) | + | + | + | + | + | + | ±* | – |
PCR results were obtained by electrophoretic analysis. LAMP results were determined visually.
+, Amplification occurred; –, amplification did not occur; ±, amplification occurred once within three trials; ±*, amplification occurred twice within three trials.