Skip to main content
. 2011 Oct;49(10):3591–3595. doi: 10.1128/JCM.00686-11

Table 3.

Comparison of sensitivity and specificity data between different commercial and in-house assays

Assaya No. of laboratories using the indicated assay EV detection
EV71 detection
CA16 detection
No. of correct positive results/total no. of positive results (% sensitivity) No. of correct negative results/total no. of negative results (% specificity) No. of correct postivie results/total no. of postive results (% sensitivity) No. of correct negative results/total no. of negative results (% specificity) No. of correct positive results/total no. of postive results (% sensitivity) No. of correct negative results/total no. of negative results (% specificity)
A 17 265/272 (97.4) 65/68 (95.6) 114/153 (74.5)b 181/187 (96.8) 146/153 (95.4) 185/187 (98.9)
B 12 191/192 (99.5) 48/48 (100) 106/108 (98.1) 131/132 (99.2) 106/108 (98.1) 132/132 (100)
C 3 44/48 (91.7) 10/12 (83.3) 24/27 (88.9) 33/33 (100) 25/27 (92.6) 33/33 (100)
D 2 28/32 (87.5)b 7/8 (87.5) 0/18 (0)b 22/22 (100) 14/18 (77.8) 21/22 (95.5)
E 1 15/16 (93.8) 4/4 (100) 7/9 (77.8) 11/11 (100) 9/9 (100) 11/11 (100)
F 1 16/16 (100) 4/4 (100) 8/9 (88.9) 11/11 (100) 9/9 (100) 11/11 (100)
In-house 5 80/80 (100) 19/20 (95) 44/45 (97.8) 53/55 (96.4) 42/45 (93.3) 55/55 (100)
Total 41 639/656 (97.4) 157/164 (95.7) 303/369 (82.1) 442/451 (98) 351/369 (95.1) 448/451 (99.3)
a

A, B, C, D, E, F, and In-house represent the same assays as described for Table 2.

b

Statistically significant result compared to in-house assay result (P < 0.0083). The criteria of test α were adjusted using the Bonferroni method (22a). α = 0.05/(group no. − 1) = 0.0083.