. 2011 Jan 31;38(2):1037–1044. doi: 10.1118/1.3544657

Table 2.

Range of errors (%) and mean absolute errors (%) for clinically relevant metrics in the case of all plans (N) meeting a specified threshold Gamma passing rate for three sets of Gamma parameters

Anatomy dose metric   Observed errorsa (%) in DVH dose metrics for plans exceeding ≥95% passing rateb (3∕3 and 2∕2 criteria) and exceeding ≥90% passing rateb (1∕1 criteria)
3%∕3 mm (N=83) 2%∕2 mm (N=51) 1%∕1 mm (N=12)
Spinal cord Range of % Errors [−11.1, 15.7] [−11.1, 15.7] [−2.7, 3.3]
D1cc Mean absolute errorc (%) 3.222 3.367 2.309
Contralateral Range of % errors [−10.9, 12.0] [−10.9, 12.0] [−5.1, 5.7]
Parotid mean Mean absolute errorc (%) 4.50 5.52 4.04
Ipsilateral Range of % errors [−3.7, 4.1] [−3.7, 4.1] [−1.4, 1.7]
Parotid mean Mean absolute errorc (%) 1.49 2.06 1.45
Larynx mean Range of % errors [−15.9, 9.2] [−7.6, 9.2] [−3.2, 3.7]
Mean absolute errorc (%) 5.66 5.32 2.50
CTV D95 Range of % errors [−3.7, 2.6] [−2.2, 2.6] [−1.6, 1.6]
Mean absolute errorc (%) 1.26 1.66 1.30
a

Error ranges and the mean absolute errors are given as percent errors (%) using the error-free plans as the baseline.

b

Analysis criteria method: Global % difference (normalized to max dose), 10% lower threshold, and γ index≤1 as the passing criterion.

c

The average of error magnitudes, i.e., absolute values of errors (%).