Product Performance Data for Selected Tests for Laboratory Diagnosis of CDI When Compared with the Results of CCCN and TC

Type and name of test Sensitivity
Specificity
Predictive value of positive result
Predictive value of negative result
CCCN TC CCCN TC CCCN TC CCCN TC
CCCN2,26,27
 Laboratory-developed cytotoxin 76 100 100 97
 TechLab (Wampole) 64–67 99 93 94
EIA for toxins2,7,28–30
 Meridian AB 96–99 48–58 94–97 95–98 51–88 69–88 99 87–92
 TechLab AB 60–96 87–99 96 92
 Remel Xpect 96 48 99 84 95 46 99 85
Lateral flow for toxins7,28,29,31
 TechLab 43–80 99 94 97 76 97
 Meridian Immunocard 96 48 99 99 95 91 99 87
GDH2,6,7,28,31,32
 Marion latex (no longer available) 68 95 59 96
 Biocite Triage 32–80 100 100 84–95
 TechLab 83–94 97 83–88 97–98
NAATs26,27,33–37
 Gen-Probe ProGastro (PCR) 85–92 77 95–96 98 69 94 98 96
 BD GeneOhm (PCR) 84–96 84–94 95 98 65–70 84–90 99 97–99
 Cepheid Xpert (PCR) 94–96 96–97 80–84 98–99
 Meridian Illumigene (LAMP) 92–98 98–99 92 99

CCCN and TC were the reference methods.

TechLab, Blacksburg, VA; Meridian Biosciences, Inc., Cincinnati, OH; Remel, Lenexa, KS; Biocite, Inc, San Diego, CA; Gen-Probe, San Diego, CA; BD GenOhm, La Jolla, CA; Cepheid, Sunnyvale, CA.

CCCN, cell culture cytotoxin neutralization; TC, toxigenic culture.