Table I.

Doses predicted based on manufacturer-provided chamber dose rates and NIST transfer dosimeter results.

Target Dose (kGy) NIST Transfer Dosimeter (kGy) Rel. % Diff.
Irradiation of samples at ambient room temp. (avg. temp. 41.5°C)
5 4 −18
25 24 −5
45 43 −5
65 63 −4
85 83 −3
Irradiation of samples in wet ice (avg. temp. 3.3°C)
5 5 −11
25 23 −7
45 39 −14
65 57 −13
85 77 −10
Irradiation of samples in dry ice (avg. temp. −80°C)
5 5 −10
25 25 −1
45 42 −7
65 65 0.2
85 81 −5