Table I.
Doses predicted based on manufacturer-provided chamber dose rates and NIST transfer dosimeter results.
| Target Dose (kGy) | NIST Transfer Dosimeter (kGy) | Rel. % Diff. |
|---|---|---|
| Irradiation of samples at ambient room temp. (avg. temp. 41.5°C) | ||
| 5 | 4 | −18 |
| 25 | 24 | −5 |
| 45 | 43 | −5 |
| 65 | 63 | −4 |
| 85 | 83 | −3 |
| Irradiation of samples in wet ice (avg. temp. 3.3°C) | ||
| 5 | 5 | −11 |
| 25 | 23 | −7 |
| 45 | 39 | −14 |
| 65 | 57 | −13 |
| 85 | 77 | −10 |
| Irradiation of samples in dry ice (avg. temp. −80°C) | ||
| 5 | 5 | −10 |
| 25 | 25 | −1 |
| 45 | 42 | −7 |
| 65 | 65 | 0.2 |
| 85 | 81 | −5 |