Table 4.

QSPECT phantom validation

No. Calibrated activity (MBq)
QSPECT activity (MBq)a
Deviation (%)a
Dead-time count loss (%) Total deviation no DTC (%)b
A B C Total A B C Total A B C Total
1 22.6 0 0 22.6 20.5 20.5 −9.5 −9.5 0.0 −9.5
2 78.7 0 0 78.7 82.1 82.1 4.3 4.3 −0.2 4.1
3 75.3 0 314 390 71.6 302 373 −5.0 –4.0 −4.2 −0.9 −5.0
4 75.2 146 314 535 65.9 131 315 512 −12.4 −10.2 0.4 −4.3 −1.3 −5.6
5 75.0 146 940 1160 63.9 131 903 1100 −14.9 −10.2 −3.9 −5.4 −2.9 −8.2
6 650 145 938 1730 616 139 891 1650 −5.2 −4.3 −5.0 −5.0 −4.4 −9.1
7 373 83.4 538 994 360 73.4 498 932 −3.4 −12.0 −7.4 −6.3 −2.5 −8.6

A and B, cylindrical insert compartments (175 cm3); C, large cylinder compartment (2,500 cm3).

aQSPECT data with dead-time correction (DTC).

bSPECT data corrected for attenuation, scatter and sensitivity, but not for dead-time.