Overall Impact on Speed |
++ |
++ |
+ |
++ |
+++ |
++ |
Primary Effect on Speed |
Lower plate heights, shorter column lengths, and higher optimum eluent velocities |
Drives fluid faster. Used in conjunction with smaller particles and narrower columns (≤ 2.1 mm i.d.) |
Improved interphase mass transfer and eddy dispersion |
Lower reduced plate height through improved eddy dispersion, improved interphase mass transfer. |
Reduces viscosity and thus pressure; allows fluid to move faster at same pressure. Improves interphase mass transfer. Can be combined with use of smaller particles |
Decreases resolution needed for analytical purposes and thus can speed up analysis considerably |
Requirements |
Needs high pressure instrument; requires dramatic decrease in extra-column broadening |
Needs high pressure instrument; requires dramatic decrease in extra-column broadening when used with small particles |
Generally needs high pressure to drive liquid through particles around 1 micron; ultra low dispersion equipment |
Requires low extra-column dispersion; almost as important as with sub-two micron porous particles |
Does not require high pressure instruments but does require specialized solvent pre-heater and column temperature controller. Generally requires narrow column (≤ 2.1 mm) to minimize thermal mismatch. No PEEK in columns when used above 100 °C |
No changes in hardware. But some approaches require extreme repeatability in peak shape and retention time. |
Equipment Cost |
High |
High |
High |
Medium |
Low |
N/A |
Primary Advantages |
Can improve speed significantly. Has minimal effect on selectivity and thus simplifies method transfer from bigger particles of same manufacture. |
Can improve speed significantly; The use of narrower columns can lower solvent consumption. |
High efficiency for separations of large molecules, especially proteins. |
Can be done with conventional HPLC equipment. |
Can be done with conventional HPLC equipment. |
No real change in instrument. |
Primary Disadvantages |
Requires mechanically very stable particles; frictional heating. |
Generally requires narrow column (≤2.1 mm i.d.) to minimize frictional heating. |
Very low sample load capacity and much less retention. Very small particles are hard to pack. |
Slightly decreased sample load capacity and retention. |
Requires thermally stable stationary phase and stable analytes. |
Mathematically complex. Must have very precise retention if deconvolution methods are used. |
Effect on relative retention |
None provided the substrate and stationary phase chemistry are the same as for larger particles (excluding pressure effects, see right). |
Small but real, and analyte dependent. |
Minimal |
Minimal under isocratic elution, larger under gradient elution. |
Considerable |
N/A |
Sample Capacity |
Good |
Good |
Much less |
Somewhat less |
Good |
N/A |
Available |
Quite a few and |
Quite a few and |
Very limited |
Few but growing. |
Quite a few |
N/A |