Table 2.

Effect of Column Length on Peak Half Width (μL)a

Column Length (mm)
k’b 30 50 100
0 2 3 4
1 4 5 7
2 6 8 11
5 12 16 22
10 22 29 41
a

Assumes column diameter of 2.1 mm, plate height of 1.8 μm.

b

Retention factor.