Table 2.

Sensitivity Measurement of Different Sensor Designs along with Percentage Improvement and Worst Case Standard Deviation from Three Different Measurements of Five Sensorsa

Measured sensitivity (nA(mg/dl)-1(mm)-2) % improvement in sensitivity % increase in surface area from Table 1 Worst case standard deviation (%) R2 values
Flat electrode 1.56 100% 100% 8% 0.9978

Square pillars

 Design 1S 2.31 148% 150% 8% 0.9923

 Design 2S 2.90 186% 200% 10% 0.9958

 Design 3S 4.45 285% 300% 8% 0.9963

Cylindrical pillars

 Design 1C 2.21 142% 139% 9% 0.9936

 Design 2C 2.73 175% 179% 8% 0.9946

 Design 3C 3.59 230% 257% 9% 0.9945
a

Response is measured at 0.5 V with respect to the Ag/AgCl (reference) electrode.