Skip to main content
. 2011 Sep 1;5(5):1137–1141. doi: 10.1177/193229681100500514

Table 2.

Sensor-Recorded Hypoglycemia and Hyperglycemia in 278 Subjects with ≥3 Months' Use of Veo System

SG (mg/dl) % of SG values, LGS off % of SG values, LGS on P valuea AUC, LGS off AUC, LGS on P valuea
<50 1.33 0.92 <0.01 0.09 0.07 <0.001

<60 3.58 2.63 0.14 0.32 0.22 0.016

<70 6.73 5.48 0.43 0.80 0.59 0.145

>180 28.02 31.34 0.05 18.13 17.50 0.029

>240 11.65 11.28 0.02 6.98 5.58 0.006

>300 4.64 3.41 <0.01 2.48 1.65 <0.001
a

P values compare LGS-on and LGS-off intervals.