Table 2.

Ranking of Pen Performance for Cartridge-Fitting Torque, Injection Force, and Dose-Setting Dial Torque

Measurement Cartridge-fitting torque (t-AUC;°•N) Injection force (N)

Verticala Obliqueb
Rankc 1 NovoPen 4 NovoPen 4 NovoPen 4 HumaPen Luxura

2 Itango ClikStar ClikStar Itango

3 ClikStar Itango HumaPen Luxura ClikStar

4 Biosulin Pen HumaPen Luxura NovoPen 4

5 HumaPen Luxura Biosulin Pen
a

Measurement not taken for Biosulin Pen due to automatic injection mechanism

b

Measurement not taken for Biosulin Pen due to automatic injection mechanism nor Itango because injection force spiked before push-button was displaced by 5 mm

c

Rankings based on numerical differences; 1 = highest ranking, 5 = lowest ranking