. 2011 Nov;18(11):1943–1950. doi: 10.1128/CVI.05122-11

Table 2.

Precision analyses for the Elecsys HBsAg II quantitative assaya

Assay Sample Mean (IU/ml) Repeatability
Intermediate precision
SD (IU/ml) CV (%) SD (IU/ml) CV (%)
Elecsys 2010 and cobas e 411 analyzers (1:100onboard dilution) HS 1 3.07 0.056 1.8 0.17 5.6
HS 2 54.3 0.747 1.4 3.04 5.6
HS 3 6,610 237 3.6 372 5.6
PreciControl HBsAg II 1 <0.05
PreciControl HBsAg II 2 0.109 0.003 3.1 0.010 8.9
Modular Analytics E170, cobas e 601, or cobas e 602 (1:400 onboard dilution) HS 4 3.06 0.069 2.3 0.15 4.9
HS 5 55.5 1.48 2.7 3.64 6.6
HS 6 37,300 1,490 4.0 3,590 9.6
PreciControl HBsAg II 1 <0.05
PreciControl HBsAg II 2 0.130 0.007 5.6 0.010 7.6
a

SD, standard deviation; CV, coefficient of variation; HS, human serum.