Representative dilution series for HBV genotypes A to Ea

Genotype Undiluted Dilution
1:100* 1:400† 1:1,000 1:4,000 1:10,000 1:40,000 1:100,000 1:1,000,000
A 123.40 >130‡ 33,052 32,640 32,120 31,800 31,240 30,300 <0.05#
B >130‡ 1,193 1,148 1,220 1,196 1,340 <0.05# <0.05# <0.05#
C >130‡ 673 728 716 744 740 <0.05# <0.05# <0.05#
D 33.31 >130‡ >130‡ 78,410 77,120 80,900 77,200 78,100 78,000
E >130‡ >130‡ 24,080 23,800 24,000 23,900 24,440 23,400 <0.05#
a

Onboard dilution for Elecsys 2010 or cobas e 411 (*) and Modular Analytics E170, cobas e 601, or cobas e 602 (), respectively. The mandatory dilution algorithm incorporating onboard dilution eliminates any potential high-dose hook effect. >130‡ indicates that the HBsAg level in the diluted sample is above the upper limit of detection for the assay; therefore, the level in the undiluted sample cannot be calculated. <0.05# indicates that the HBsAg level in the diluted sample is below the lower limit of detection for the assay; therefore, the level in the undiluted sample cannot be calculated.