Table 2.
Clinical sensitivities and mean CT values for real-time PCR kitsa
Kit | No. of positive specimens identified/total no. of positive specimens | Sensitivity (%) | Specificity (%) | Mean CT ± SD | P valued |
---|---|---|---|---|---|
In-house kit | 66/66 | 100 | 100 | 25.5 ± 8.6 | |
Argene kit | 60/66 | 90.9 | 100 | 27.1 ± 9.5 | NS |
Cepheid kit | 62/66 | 93.9 | 100 | 25.8 ± 9.2 | NS |
Focus Diagnostics kit | 64/66 | 97 | 100 | 24.8 ± 8.8b | NS |
Shanghai ZJ Bio-Tech kit | 34/66 | 51.5 | 100 | 37.3 ± 8c | <0.01 |
Data for the four real-time PCR kits tested are in comparison to results of the in-house real-time PCR assay for detection of the Bordetella IS481 sequence in 66 Bordetella-positive specimens.
As recommended by the manufacturer, the CT retained for the comparison was 15.
As recommended by the manufacturer, the CT retained for the comparison was 7.
Determined by paired t test (2). NS, not significant.