. 2011 Nov;49(11):3943–3946. doi: 10.1128/JCM.00335-11

Table 2.

Clinical sensitivities and mean CT values for real-time PCR kitsa

Kit No. of positive specimens identified/total no. of positive specimens Sensitivity (%) Specificity (%) Mean CT ± SD P valued
In-house kit 66/66 100 100 25.5 ± 8.6
Argene kit 60/66 90.9 100 27.1 ± 9.5 NS
Cepheid kit 62/66 93.9 100 25.8 ± 9.2 NS
Focus Diagnostics kit 64/66 97 100 24.8 ± 8.8b NS
Shanghai ZJ Bio-Tech kit 34/66 51.5 100 37.3 ± 8c <0.01
a

Data for the four real-time PCR kits tested are in comparison to results of the in-house real-time PCR assay for detection of the Bordetella IS481 sequence in 66 Bordetella-positive specimens.

b

As recommended by the manufacturer, the CT retained for the comparison was 15.

c

As recommended by the manufacturer, the CT retained for the comparison was 7.

d

Determined by paired t test (2). NS, not significant.