. 2004 Jan;11(1):98–101. doi: 10.1128/CDLI.11.1.98-101.2004

TABLE 1.

Sensitivity and specificity of the Determine Rapid Syphilis TP assay by reader type with the TPHA as the reference standarda

Reader TPHA positive (n = 250)
TPHA negative (n = 300)
No. determine positive % Sensitivity (95% CIb) No. determine negative % Specificity (95% CI)
Laboratorian 1c 244 97.6 (94.6-98.8) 292 97.3 (94.6-98.6)
Laboratorian 2c 246 98.4 (95.6-99.4) 287 95.7 (92.9-97.7)
Nursed 241 95.6 (93.2-96.2) 289 96.3 (93.4-98.3)
a

Five hundred fifty samples were tested; 17 TPHA-indeterminate results are not included.

b

CI, confidence interval.

c

All TPHA-indeterminate results were read as negative.

d

Fifteen of 17 indeterminate TPHA results were read as negative.