TABLE 3.

Sensitivity and specificity of the Determine Rapid Syphilis TP assay by reader type and HIV status using the TPHA as the reference standarda

Reader HIV positive (n = 198)
HIV negative (n = 127)
% Sensitivity (95% CIb) % Specificity (95% CI) % Sensitivity (95% CI) % Specificity (95% CI)
Laboratorian 1 98.5 (93.7-99.6) 95.5 (86.3-98.5) 94.4 (83.3-98.2) 100 (88.6-100%)
Laboratorian 2 99.2 (94.2-99.8) 92.5 (82.5-96.8) 96.3 (85.2-99.0) 97.3 (88.8-99.3)
Nurse 97.7 (92.4-99.2) 92.5 (82.2-98.9) 94.4 (83.1-98.1) 98.6 (89.8-99.8)
a

Three hundred twenty-five samples were tested, excluding samples without clinical data (n = 227) and those TPHA indeterminate (n = 15).

b

CI, confidence interval.