Table 1.

Parameters for adiabatic RF pulses used in simulations and experiments

Sequence block RF pulse B1 modulation G modulation(a) Tp
(ms)
BW
(kHz)
γB1,max
(kHz)
TOCSY (high-power) GOIA-W(16,4) WURST-16 WURST-4 1.25 16 1.22
TOCSY (medium-power) GOIA-W(16,4) WURST-16 WURST-4 1.5 13.34 1.02
TOCSY (low-power) (b) GOIA-W(16,4) WURST-16 WURST-4 2.0 10 0.76
LASER GOIA-W(16,4) WURST-16 WURST-4 3.5 20 0.82
a)

for all pulses an inverse WURST-4 function was used with a gradient reduction factor f = 0.9 in the middle of the pulse;

b)

the condition used for in-vivo MRS.