. 2010 Oct 8;81(5):526–534. doi: 10.3109/17453674.2010.519169

Table 2.

Wear rates of MOM bearing couples defined in different units

Type of wear Wear rate Method Ref.
Linear wear rate of femoral heads per year 7.6 μm (range 2.9–13) to Explanted implant(s) Reinisch et al. (2003)
250 μm (range 50–810) Radiographic analysis Stilling et al. (2009)
 First year 25 μm Explanted implant(s) Sieber et al. (1999)
 > 3 years 5 μm Explanted implant(s) Sieber et al. (1999
Volumetric wear rate of femoral heads per year 2.0 mm3 (range 0.55–3.7) Explanted implant(s) Reinisch et al. (2003)
5.0 mm3 (range 0.22–22) Explanted implant(s) Willert et al. (1996)
Mass wear rate per year 17 mg (range 4.6–31) Explanted implant(s) Reinisch et al. (2003)
No. of particles per unit volume of wear per mm3 2.7 × 1012 – 1.5 × 1013 Pin-on-plate Tipper et al. (1999)
Number of particles per 106 cycles 4 × 1012 – 6 × 1013 Pin-on-plate Tipper et al. (1999)
Number of particles per year 6.7 × 1012 – 2.5 × 1014 Explanted implant(s) Doorn et al. (1998)

Note: Retrieval study data were obtained from patients undergoing revision of THAs with MOM bearing couples. Radiographic wear analysis was performed by analysis of digitized anteroposterior (AP) radiographs using a computerized method.