Table 2.
Analysis method | Mean (range) (mm) |
SDdif-intr a | Bias b (±LOA) c(mm) | 95% CI d(mm) | p-value e |
---|---|---|---|---|---|
2D measurements | |||||
PW2 f | 0.80 (0.28–1.78) | 0.26 | -0.09 (±0.53) | -0.26–0.08 | 0.25 |
PW1 g | 1.07 (0.69–1.47) | 0.11 | 0.04 (±0.22) | -0.04–0.11 | 0.29 |
EGS-RSA h | 0.60 (0.13–1.09) | 0.11 | 0.06 (±0.23) | -0.02–0.13 | 0.11 |
3D measurements | |||||
PW2 | 1.12 (0.27–2.20) | 0.31 | -0.05 (±0.62) | -0.25–0.15 | 0.61 |
PW1 | 1.48 (0.86–2.31) | 0.44 | -0.03 (±0.87) | -0.31–0.25 | 0.82 |
EGS-RSA | 0.75 (0.26–1.47) | 0.16 | 0.05 (±0.31) | -0.05–0.15 | 0.33 |
a SDdif-intra is the random variation within a method comparing double examinations.
b Bias: systematic variation within a method.
c LOA: limits of agreement around the bias (95% prediction interval = SDdif-intra x 1.96).
d 95% confidence interval for the bias.
e p value (paired t-test) bias between methods.
f PW2: PolyWare PE wear analysis using the postoperative and final follow-up radiographs.
g PW1: PolyWare PE wear analysis using only the final radiographic follow-up radiographs.
h EGS-RSA: radiostereometric analysis of PE wear using sphere models (the “gold standard”).