. 2010 Oct 8;81(5):570–578. doi: 10.3109/17453674.2010.506632

Table 2.

Repeatability of radiographic double-examination wear measurements for the 3 methods

Analysis method Mean (range)
(mm)
SDdif-intr a Bias b (±LOA) c(mm) 95% CI d(mm) p-value e
2D measurements
 PW2 f 0.80 (0.28–1.78) 0.26 -0.09 (±0.53) -0.26–0.08 0.25
 PW1 g 1.07 (0.69–1.47) 0.11 0.04 (±0.22) -0.04–0.11 0.29
 EGS-RSA h 0.60 (0.13–1.09) 0.11 0.06 (±0.23) -0.02–0.13 0.11
3D measurements
 PW2 1.12 (0.27–2.20) 0.31 -0.05 (±0.62) -0.25–0.15 0.61
 PW1 1.48 (0.86–2.31) 0.44 -0.03 (±0.87) -0.31–0.25 0.82
 EGS-RSA 0.75 (0.26–1.47) 0.16 0.05 (±0.31) -0.05–0.15 0.33

a SDdif-intra is the random variation within a method comparing double examinations.

b Bias: systematic variation within a method.

c LOA: limits of agreement around the bias (95% prediction interval = SDdif-intra x 1.96).

d 95% confidence interval for the bias.

e p value (paired t-test) bias between methods.

f PW2: PolyWare PE wear analysis using the postoperative and final follow-up radiographs.

g PW1: PolyWare PE wear analysis using only the final radiographic follow-up radiographs.

h EGS-RSA: radiostereometric analysis of PE wear using sphere models (the “gold standard”).