Comparison of the image contrasts measured on the contrast rods of the Catphan©600 phantom.

  ROI 1 2 3 4 5 6 7
Contrast CBCT with correction 137 190 271 1061 874 249 1058
  CBCT without correction 69 97 141 557 414 121 562