Skip to main content
. 2004 Jan;42(1):16–20. doi: 10.1128/JCM.42.1.16-20.2004

TABLE 4.

Sensitivity and specificity of the Amplicor DNA test on DBSa

PBMC nested PCR resulta (no. of samples) Amplicor DNA test v1.5 on DBS (no. of samples)
First testb
Repeatc
+ Ind + Ind
+ (10) 10 0 0 10 0 0
− (117) 9 106 2 0 115 2
Ind (23) 0 23 0 0 23 0
a

+, HIV-positive result; −, HIV-negative result; Ind, inderminate result for HIV.

b

Sensitivity, 100%; specificity, 90%.

c

Sensitivity, 100%; specificity, 98%.