AxSYM Combo precision

Proficiency panel member No. of replicates Mean S/CO Precision
Within runs
Between runs
SD CV (%) SD CV (%)
Index calibrator 189 17.10a 0.824 4.8 0.972 5.7
AxSYM Combo negative assay control 414 0.38 0.017 4.5 0.021 5.5
AxSYM Combo HIV-1-positive assay control 414 3.84 0.205 5.3 0.253 6.6
AxSYM Combo HIV-2 positive assay control 414 3.55 0.166 4.7 0.219 6.2
AxSYM Combo viral lysate positive assay control 414 3.23 0.158 4.9 0.191 5.9
Anti HIV-1 high titer (nondiluted) 189 25.96 1.29 5.0 2.41 9.3
Anti HIV-2 high titer (nondiluted) 189 54.15 2.00 3.7 2.47 4.6
Anti HIV-1 medium titer (diluted) 189 11.85 0.541 4.6 0.742 6.3
Anti HIV-2 medium titer (diluted) 189 11.79 0.519 4.4 0.912 7.7
Anti HIV-1 group O 189 3.92 0.283 7.2 0.333 8.5
HIV p24 antigenb 162 1.59 0.068 4.2 0.102 6.4