Published in final edited form as: Circ Cardiovasc Imaging. 2011 Sep 23;4(6):662–670. doi: 10.1161/CIRCIMAGING.111.965764

Table 2.

Predictor of Cardiac Function Evaluation by Cine CMR

Cardiac Function Evaluation by Cine CMR in patients with Left-sided ICD/BiV-ICD
P-Value
Possible
N=47 Scans
Impossible
N=8 Scans
Age 58±14 64±20 0.27
Female/Male 7/39 1/7 0.84
Body Mass Index [kg/m2] 26.9±3.9 23.1±2.8 0.019*
Ejection Fraction [%] 34.0±16.1 34.0±17.1 0.996
Left Ventricular End-Diastolic Diameter [mm] 55.3±10.8 65.0±7.7 0.045*
Artifact Size on Short Axis of Cine MRI [cm2] 200.3±32.4 186.6±19.7 0.478
Minimum Distance from Generator to Heart [mm]
 Frontal Chest X-ray 31.7±20.9 15.7±15.0 0.08
 Lateral Chest X-ray 32.2±4.0 33.2±3.6 0.58
Generator
 Height*Width [cm2] 38.7±6.4 39.2±7.5 0.82
 Thickness [mm] 13.0±1.8 13.0±1.7 0.99
 Weight [g] 80.6±8.3 81.5±7.2 0.78
 Volume [cc] 36.1±5.9 35.1±5.0 0.65

Values are shown as mean ± SD or N.

Significant P-Value defined as P<0.05 are shown by the asterisk (*).