Effect of interfering substances on the performance of the HA ELISA

Interfering substance & conc. tested Specimen type Conc. Unspiked (IU/ mL) Conc. Spiked (IU/ mL) Deviation observed (%)
Hemoglobin (486 mg/dL) Negative 8.8 12.0 36.4
Low positive 61.6 60.1 2.4
High positive 719.6 760.0 5.6

Bilirubin (20.3 mg/dL) Negative 8.8 5.9 33.0
Low positive 61.6 48.5 21.3
High positive 719.6 653.3 9.2

 Lipemia*
  2000 mg/dL Negative 8.8 7.9 10.2
Low positive 61.6 49.4 19.8
High positive 719.6 703.2 2.3
  1000 mg/dL Low positive 61.6 62.3 1.1
High positive 719.6 710.1 1.3
*

Evaluation of lipemia on the analytic performance of the HA ELISA was based on our in-house protocol as the exact conversion from 1460 Units suggested by the manufacturer to mg/dL could not be ascertained.