Published in final edited form as: Acta Biomater. 2011 Sep 1;8(1):133–143. doi: 10.1016/j.actbio.2011.08.023

Table 3.

Effect of Autoclave Sterilization on Thermal Properties

Polymer Compositiona Tg (°C) Tm (°C) Tc (°C) ΔHm J/g ΔHc J/g Percent Crystallinity
PCLFDEG −57.2
(−1.4)
39.4
(2.1)
13.0
(16.3)
56.3
(15.6)
53.6
(14.2)
41.7
(11.6)
PCLFPPD100 −58.2
(−1.4)
37.8
(3.5)
9.3
(8.2)
44.9
(4.6)
48.3
(11.3)
33.2
(4.1)
PCLFPPD75PCLFGLY25 −58.6
(−0.9)
36.7
(2.4)
9.7
(8.7)
44.7
(7.0)
45.0
(10.7)
33.1
(5.2)
PCLFPPD50PCLFGLY50 −57.5
(−0.4)
33.9
(1.8)
5.2
(8.8)
38.1
(3.1)
39.3
(3.1)
28.2
(2.3)
PCLFPPD25PCLFGLY75 −57.1
(−0.4)
32.0
(2.1)
5.2
(11.5)
43.1
(11.0)
45.0
(10.1)
32.0
(8.2)
PCLFGLY100 −56.8
(−0.5)
24
(1.5)
−8.2
(4.7)
34.1
(3.4)
37.6
(8.0)
25.3
(2.5)

Thermal characterization of cross-linked PCLF scaffolds after autoclave sterilization. The values measured are shown on top, and the difference between the values measured before and after autoclave sterilization are in ( ).