Step # |
Volume (mL) |
Description |
Parameters |
1 |
0.00 |
Collect 1.0 mL fractions during entire run |
|
|
2 |
0.00 |
Switch Columns |
HIC Column 1
(Position 2) |
|
3 |
0.00 |
Isocratic Flow |
pH: 6.80
100% B |
Volume: 2.00 mL
Flow: 1.00 mL/min |
4 |
2.00 |
Zero Baseline |
QuadTec |
|
5 |
2.00 |
Zero Baseline |
UV Detector |
|
6 |
2.00 |
Inject Sample |
Sample Load
Dynamic Loop |
Auto Inject Valve
Volume: 0.50 mL
Flow: 1.00 mL/min |
7 |
3.00 |
Isocratic Flow |
pH: 6.80
100%B |
Volume: 5.00 mL
Flow: 1.00 mL/min |
8 |
8.00 |
Linear Gradient |
pH: 6.80
100%B -> 0%B |
Volume: 10.00 mL
Flow: 1.00 mL/min |
9 |
18.00 |
Isocratic Flow |
pH: 6.80
0%B |
Volume: 3.00 mL
Flow: 1.00 mL/min |
10 |
21.00 |
Isocratic Flow |
pH: 6.80
100%B |
Volume: 5.00 mL
Flow: 1.00 mL/min |
11 |
26.00 |
Switch Columns |
Union
(Position 1) |
|
End |
26.00 |
End of Protocol |
Automatically Repeats with 5 additional HIC columns
(Positions 3-7) |
Scout type |
Number of Runs: 6 |
Number of Steps Scouted: 1 |